• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/86
Device Problem Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/24/2023
Event Type  malfunction  
Event Description
It was reported that the during lead implant procedure, the guidewire failed to be removed from the novel left ventricular lead.The procedure was completed using an alternate lead on (b)(6) 2023.The patient was stable.
 
Manufacturer Narrative
The reported event of stylet ¿unable to remove¿ was confirmed.As received, a complete lead was returned in one piece with the stylet stuck inside the lead.Final analysis found the coating of the stylet stripped and bunched with the inner coil at the connector region.The cause of the reported event was isolated to bunching of the stripped stylet coating and inner coil.No other anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18104068
MDR Text Key327773300
Report Number2017865-2023-52012
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberA000128617
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/21/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age71 YR
Patient SexMale
Patient Weight107 KG
-
-