• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Catalog Number CA13E0/300/000JP
Device Problems Break (1069); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/01/2023
Event Type  malfunction  
Manufacturer Narrative
H4: device manufacture date is unknown, no information has been provided to date.G5: 510k is blank, device is exempt.One device was received.Per visual inspection, the gas sampling port's connector was caught in the hme filter.No other anomaly was observed.A device history record (dhr) review could not be performed as the lot number was unknown.The returned device has been forwarded to the manufacturing site for a detailed investigation.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that during a pre-use check.The gas sampling line connection was damaged.No patient injury or clinical affects was reported.
 
Manufacturer Narrative
Other text: d3, g1, and g2 email is: regulatory.Responses@icumed.Com.H6 - evaluation codes updated.Device evaluation: the product was forwarded to the secondary investigation site.It was determined that part number was manufactured by a supplier.Complaint investigation was completed by the supplier.The supplier had never seen the item broken in that are previously.They determined the broken area "suffered a very big force.Which may have come from transportation, handling, or the opening method is twisted instead of pulling out".Cause of damage could not be fully established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key18106950
MDR Text Key329199934
Report Number3012307300-2023-10420
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCA13E0/300/000JP
Was Device Available for Evaluation? No
Date Returned to Manufacturer10/23/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/15/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-