It was reported that the health care professional (hcp) called to report that this cardiac resynchronization therapy defibrillator (crt-d) recorded a high out of range impendence on the left ventricular (lv) lead.The patient was brought into the clinic for an in person visit and lead measurements were taken came back normal and within range.However, another high out of range impendence alert was recorded after the visit.The hcp called technical services (ts) to review the stored data.Ts reviewed the presented electrogram (egm) and lv loss of capture (loc) was noted.Ts recommended for further lead testing be performed.The lv lead remains in service.No adverse patient effects were reported.Subsequent additional information was received reported that this left ventricular (lv) lead exhibited high pacing threshold, high out of range impedance measurements and loss of capture (loc).A chest xray imaging was done indicated possible fracture.The physician performed a lead revision procedure, where the lv lead was surgically abandoned and the crtd device was explanted.The crtd and lv lead were successfully replaced with a new device and lead.No additional adverse patient effects were reported.
|