(b)(4).Method: the complaint mr290v vented autofeed humidification chamber was returned to f&p healthcare in new zealand where it was visually inspected and analysed.Results: visual inspection of the returned mr290v chamber confirmed three cracks.Two vertical crack lines were situated to the left and right of the chamber bracket.A third crack originated at water level marking below one of the ports and extended vertically toward the base where it stretches horizontally to the left and right alongside the base.No damage was found to the top of the dome or the chamber base.Conclusion: we are unable to determine the cause of the damage to the humidification chamber.The mr290v chambers are designed and tested to conform to iso 5367 breathing tubes intended for use with anaesthetic apparatus and ventilators.Every mr290v chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber dome due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.No cracks in the chamber dome are acceptable.Any chamber that fails this inspection is rejected.Our user instructions that accompany the mr290v vented autofeed humidification chamber state the following: - "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." - "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." - "use of the mr290 above the maximum operating pressure may lead to cracking, water leakage and, on rare occasions, could lead to a loss of ventilation pressure." - "do not use the chamber if the seals are not intact when received, or if it has been dropped.".
|