Model Number VLS-1190STK
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/26/2023
Event Type  malfunction  
Manufacturer Narrative
This device is class1 product so that 510k# is blank.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the distal end with ccd module as defective.Based on the result, we concluded that it was caused due to the excessive force applied on the distal end with ccd module.In addition, our technician confirmed that the insertion flexible tube fluid damage, the control body fluid damage, the ccd drive pcb fluid damage, the control body corroded, the lg connector corroded, the bending rubber missing, and the remote control buttons cut; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Event Description
The time of event is unknown.There was no report of patient harm.Video image failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO LARYNGOSTROBOSCOPE (90 SERIES)
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315800
MDR Report Key18108334
MDR Text Key328501780
Report Number9610877-2023-60864
Device Sequence Number1
Product Code EQL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 11/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVLS-1190STK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/26/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-