LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 10-85-00 |
Device Problems
Inaccurate Flow Rate (1249); Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/13/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 double head pump.The incident occurred in oyama, japan.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that the flow rate adjustment knob on the side a of a s5 double head pump came off and was rattling during procedure.There was no patient injury.
|
|
Manufacturer Narrative
|
H10: a livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue: flow rate adjustment knob itself can come loose from its fixation.To solve the issue, knob screw was fixed.Subsequent functional verification testing was completed without further issues and the unit was returned to service.Most likely root cause of the reported event was knob looseness during use.Based on the information collected and investigation carried out, since the involved pump is a roller one, its shaft encoder movements are limited by mechanical constraint of the housing and pump speed can be anyway controlled by the user.Therefore, based on this rationale the reported event (loose roller pump knob) is not likely to result in serious injury and has been re-assessed as not reportable.
|
|
Event Description
|
See initial report.
|
|