MEDTRONIC HEART VALVES DIVISION EVOLUT PRO PLUS VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV Back to Search Results
Model Number EVPROPLUS-34US
Device Problems Fluid/Blood Leak (1250); Patient Device Interaction Problem (4001); Material Split, Cut or Torn (4008); Central Regurgitation (4068)
Patient Problems Cusp Tear (2656); Heart Failure/Congestive Heart Failure (4446); Aortic Valve Insufficiency/ Regurgitation (4450)
Event Date 11/01/2023
Event Type  Injury  
Manufacturer Narrative
Product analysis: the device remains implanted, and no procedural images were submitted for review; therefore, no product analysis can be performed.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Medtronic has requested additional information pertaining to this reportable event.If additional reportable information is received, a supplemental report will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that three years and nine months following the implant of this transcatheter bioprosthetic valve, the valve was failing due to a torn leaflet causing severe aortic insufficiency.Subsequently, the patient went into heart failure.A second non-medtronic valve was implanted successfully, and the patient was doing well.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received that severe central aortic insufficiency was noted.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT PRO PLUS VALVE
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18110387
MDR Text Key327881625
Report Number2025587-2023-04669
Device Sequence Number1
Product Code NPT
UDI-Device Identifier00763000211158
UDI-Public00763000211158
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/23/2020
Device Model NumberEVPROPLUS-34US
Device Catalogue NumberEVPROPLUS-34US
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/06/2023
Initial Date FDA Received11/10/2023
Supplement Dates Manufacturer Received11/10/2023
Supplement Dates FDA Received11/17/2023
Date Device Manufactured10/04/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age73 YR
Patient SexMale
-
-