SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUIT, EXPANDABLE; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Catalog Number C45101344-NLJ
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/01/2023
Event Type  malfunction  
Manufacturer Narrative
Other, other text: b3: month and year of event have been provided, day is unknown.D4: udi section is unknown, no information has been provided to date.G5: 510k is blank, device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that during pre-testing, there was a crack in the circuit connection.No adverse effects have been reported.
 
Manufacturer Narrative
One device and three photos were returned for investigation.Per visual inspection of the device, a broken connector was found.The reported complaint was confirmed, and this problem causes an air leak in the circuit.No corrective action was taken.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ANESTHESIA BREATHING CIRCUIT, EXPANDABLE
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key18110567
MDR Text Key327931698
Report Number3012307300-2023-10431
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC45101344-NLJ
Device Lot Number4398591
Was Device Available for Evaluation? No
Date Returned to Manufacturer10/25/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/24/2023
Initial Date FDA Received11/10/2023
Supplement Dates Manufacturer Received01/23/2024
Supplement Dates FDA Received02/21/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/12/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-