• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF UNIT "ESG-400"; ELECTROSURGICAL GENERATOR ESG-400

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF UNIT "ESG-400"; ELECTROSURGICAL GENERATOR ESG-400 Back to Search Results
Model Number WB91051W
Device Problems Communication or Transmission Problem (2896); Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned for evaluation.If additional information becomes available prior to the conclusion of the investigation, a supplemental report will be filed.
 
Event Description
The customer reported that his olympus hf generator unit was resetting continuously and producing an e433 error code.Additional details relating to the patient and the event have been requested, but no response has been received at this time.There was no patient harm or consequence reported as a result of this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's device evaluation and investigation.Based on the results of the device evaluation, the reported event was confirmed along with burn traces.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 9 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the reported event could not be identified.However, it¿s likely the cause is related to the ingress of moisture which led to damage/short circuitry of the generator board and motherboard.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF UNIT "ESG-400"
Type of Device
ELECTROSURGICAL GENERATOR ESG-400
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18111144
MDR Text Key329223208
Report Number9610773-2023-03239
Device Sequence Number1
Product Code GEI
UDI-Device Identifier04042761076838
UDI-Public04042761076838
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K203682
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWB91051W
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/06/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-