Catalog Number EMC9608 |
Device Problems
Break (1069); Fluid/Blood Leak (1250)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 10/05/2023 |
Event Type
Injury
|
Event Description
|
It was reported that during an intravenous (iv) infusion with a solution administration set, the patient "sat up and pulled something found bleeding".It was further reported that "on inspection", the set was noted to be damaged ("snapped"), and blood from the central venous catheter back tracked onto bed.There was no report of medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
The actual device was not available for evaluation; however, two retention samples were evaluated.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.All junctions underwent a push-pull test, and no disconnections were identified.The samples were leak tested underwater and air passage testing was performed.No leaks were identified, and no blockages were found.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|