BAXTER HEALTHCARE CORPORATION SOLUTION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number EMC9608
Device Problems Break (1069); Fluid/Blood Leak (1250)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 10/05/2023
Event Type  Injury  
Event Description
It was reported that during an intravenous (iv) infusion with a solution administration set, the patient "sat up and pulled something found bleeding".It was further reported that "on inspection", the set was noted to be damaged ("snapped"), and blood from the central venous catheter back tracked onto bed.There was no report of medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The actual device was not available for evaluation; however, two retention samples were evaluated.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.All junctions underwent a push-pull test, and no disconnections were identified.The samples were leak tested underwater and air passage testing was performed.No leaks were identified, and no blockages were found.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLUTION SETS
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MALTA
a47 industrial estate
marsa
MT  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18113834
MDR Text Key327878346
Report Number1416980-2023-05841
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEMC9608
Device Lot Number23F14V846
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-