Through implant patient registry it was learned that a 27mm 8300ab valve in the aortic position, was explanted after an implant duration of 5 years, 3 months due to calcification and pannus (scarred to lvot).The patient presented with sob and fatigue.The explanted valve was replaced with a 29mm 11060a valved conduit.The patient was in stable condition post procedure and was discharged on pod#10.The patient is a 75 year old male with history of hld, htn, cad.
|
The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Calcific degeneration is a common cause of bioprosthetic heart valve failures.Many factors contribute to the onset and propagation of calcification.These include patient factors (age, disease state, pharmacological intervention, etc.), mechanical stress related to the valves hemodynamic performance, and glutaraldehyde fixation of tissue.Of these, the fixation process is a relatively minor contributor to calcification for edwards tissue valves due to anti-calcification treatments during manufacturing.Though numerous studies have been conducted on preventive calcification strategies in bioprosthetic heart valves, the causes of calcification are not fully understood and there are still no mechanisms or medical therapies which fully prevent bioprostheses from calcifying.Calcification is most commonly related to patient factors and is not usually an indication of a device malfunction related to a manufacturing deficiency.The instructions for use (ifu) have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.The most likely cause is patient factors, including hyperlipidemia (hld) and coronary artery disease.Pannus overgrowth, or host tissue, is considered to be a form of non-structural valve dysfunction.The growth of host tissue on the sewing ring is expected and is a natural part of the healing reaction to prosthesis implantation.A small amount of host tissue growth over the suture line is needed to form a non-thrombogenic surface and complete the healing process after valve implantation.In contrast, if there is an excessive amount of pannus growth, it can extend onto the cusp surfaces leading to thickening of the cusps, leaflet immobility, elevated gradients, and stenosis.Host tissue growth can also contribute to cusp retraction or curling resulting in valvular regurgitation.Host fibrous (pannus) tissue growth is not a malfunction of the device related to a manufacturing deficiency.The instructions for use (ifu) have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.The most likely cause is patient factors.
|