Model Number CD3371-40QC
Device Problem Incorrect Measurement (1383)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2023
Event Type  malfunction  
Event Description
Related manufacture reference number: 2017865-2023-52161.It was reported that the patient presented during clinical follow-up.Interrogation noted overestimation of battery longevity.No interventions were performed.There were no patient consequences.
 
Manufacturer Narrative
The reported event of overestimation could not be confirmed.The results of the investigation are inconclusive since the product was not returned for analysis.Device session records were provided and software investigation was performed.Review of information provided confirmed that the longevity calculation was correct based on the variation of battery performance and the fuel gauge/longevity calculation.The device is performing as expected and no anomalies were observed.
 
Manufacturer Narrative
G3 correction: the statement of the previous follow-up report was incorrect.The awareness date of the initial report is 19 oct 2023.The awareness date of the first follow-up report is should have been 5 feb 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18117432
MDR Text Key327882888
Report Number2017865-2023-51959
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 02/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2019
Device Model NumberCD3371-40QC
Device Lot NumberP000049060
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/12/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/18/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial