Catalog Number 32-422703 |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/16/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the oxford trail bearing fractured during surgery.This event is related to malfunction that could potentially lead to serious injury.Due diligence is in progress for this complaint; to date no additional information or product has been received.
|
|
Manufacturer Narrative
|
(b)(4).G2- foreign- canada.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.Visual examination of the returned product confirmed that the item and lot numbers match that which is on the complaint.The instrument looked well used and featured multiple dents and scuffs to its surfaces.It was confirmed that the instrument exhibited a fracture in the area of the item number and the broken off section of the bearing was not returned for evaluation.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Device is used for treatment.Medical records were not provided.With the available information, a definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Search Alerts/Recalls
|