Model Number 71020
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Inadequate Pain Relief (2388)
Event Date 10/19/2023
Event Type  Injury  
Event Description
Patient was implanted with the nalu peripheral nerve stimulator on (b)(6) 2022 to treat lower back pain.Approximately one year after being implanted and after significant weight loss, the patient reported having issues with communication between the implantable pulse generator (ipg) and the external therapy discs.Assessment by the physician determined that the patient's loose skin resulting from the weight loss was interfering with the device communication.Surgical revision was performed on (b)(6) 2023 to place a new ipg and leads in a more medial location to avoid the excess skin and improve connectivity.
 
Manufacturer Narrative
There are no allegations of device or component failure or malfunction and no evidence of any component migration.The communication issues reported appear to be directly related to the change in patient condition and were resolved by moving the system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NALU NEUROSTIMULATION SYSTEM
Type of Device
PERIPHERAL NERVE STIMULATOR
Manufacturer (Section D)
NALU MEDICAL INC
2320 faraday avenue
suite 100
calsbad CA 92008
Manufacturer (Section G)
NALU MEDICAL, INC
2320 faraday ave
suite 100
carlsbad
Manufacturer Contact
terry villarba
2320 faraday avenue
suite 100
carlsbad, CA 92008-7241
7604482360
MDR Report Key18126963
MDR Text Key327997304
Report Number3015425075-2023-00281
Device Sequence Number1
Product Code GZF
UDI-Device Identifier00812537036182
UDI-Public01008125370361821122033017250330
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183579
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71020
Device Catalogue Number71020
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/30/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-