• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WARSAW ORTHOPEDICS PRESTIGE CERVICAL DISC SYSTEM; PROSTHESIS, INTERVERTEBRAL DISC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WARSAW ORTHOPEDICS PRESTIGE CERVICAL DISC SYSTEM; PROSTHESIS, INTERVERTEBRAL DISC Back to Search Results
Model Number 6972860
Device Problem Migration (4003)
Patient Problem Unspecified Infection (1930)
Event Date 10/20/2023
Event Type  Injury  
Event Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having device used for cervical fusion.Levels: c5/6 it was reported that cage settling occurred.The cage settled a couple millimeters into the end plate of the vertebral bodies.Due to this, the surgeon performed a acdf using cage and plate.Additional surgery was performed to explant the cage using distraction and kocher clamp.Patient had infection.There were no further complications reported regarding the event.
 
Manufacturer Narrative
Radiographic image review: lateral x-ray c5-6 artifical disc above c6-7 acdf possible lucency of c6-7 between graft and c6 ap view shows slight lateral displacement of superior c5-6 disc protrusion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESTIGE CERVICAL DISC SYSTEM
Type of Device
PROSTHESIS, INTERVERTEBRAL DISC
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer (Section G)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key18128725
MDR Text Key327991495
Report Number1030489-2023-00762
Device Sequence Number1
Product Code MJO
UDI-Device Identifier00613994493637
UDI-Public00613994493637
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P090029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6972860
Device Catalogue Number6972860
Device Lot Number0799464W
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/20/2023
Initial Date FDA Received11/14/2023
Date Device Manufactured08/25/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age65 YR
Patient SexMale
Patient Weight113 KG
-
-