MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VH-4000
Device Problem Intermittent Continuity (1121)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/17/2023
Event Type  malfunction  
Event Description
The hospital reported that during an endoscopic vein harvesting procedure, power to the vasoview hemopro 2 was intermittent.Another kit was opened to complete the procedure.There was no patient injury.
 
Manufacturer Narrative
Trackwise (b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure, power to the vasoview hemopro 2 was intermittent.It's not known if the polyfuse was in effect.There's no indicator of that activation on the device.Another kit was opened to complete the procedure.There was no patient injury.
 
Manufacturer Narrative
Trackwise#: (b)(4).Updated sections: a2, a3, a4, b3, b4, b5, g4, g7, h2, h3, h6, h10 a lot history record review was completed for lots 3000339321, 3000339323, and 3000337271 the last 3 lots shipped to the account prior to the event/aware date.There were no ncmrs, rework, or deviations documented for the last 3 lots shipped to the account.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.H3 other text : device discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO 2
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key18130604
MDR Text Key328038882
Report Number2242352-2023-00932
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00607567700406
UDI-Public00607567700406
Combination Product (y/n)N
PMA/PMN Number
K101274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician Assistant
Type of Report Initial,Followup
Report Date 11/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/14/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVH-4000
Device Catalogue NumberVH-4000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/20/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN.
Patient Age62 YR
Patient SexMale