MEDTRONIC PUERTO RICO OPERATIONS CO. VIVA¿ QUAD XT CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
|
Back to Search Results |
|
Model Number DTBA1QQ |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problems
Headache (1880); Nausea (1970); Diaphoresis (2452)
|
Event Date 10/06/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
It was reported that the patient was experiencing symptoms similar to ventricular tachycardia (vt), but the cardiac resynchronization therapy defibrillator (crt-d) had not recorded any vt episodes.It was noted that the patient experienced a flushed feeling, sweats, heaviness in their face, nausea, and a headache.It was also noted that there was a decrease in biventricular pacing.The crt-d was reprogrammed and remains in use. no further patient complications have been reported as a result of this event.
|
|
Search Alerts/Recalls
|
|
|