Model Number 23786 |
Device Problem
Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/1901 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional manufacuring narrative: other, other text: the device has not been returned to the investigation facility to allow for an analysis to be performed.If the product is received for evaluation or new information is obtained, a follow up report will be submitted.E1.Initial reporter phone number exceeded allowable field length, initial reporter phone number is as follows: (b)(6).
|
|
Event Description
|
The customer reported the radical-7 device "switches off automatically after about 1 minute." there was no patient impact or consequence reported.
|
|
Event Description
|
The customer reported the radical-7 device "switches off automatically after about 1 minute." there was no patient impact or consequence reported.
|
|
Manufacturer Narrative
|
Additional manufacuring narrative: other, other text: the returned device was evaluated.The radical-7 issued a 10 beep alarm prior to shut down.The reported issue was a result of a damaged component on the instrument board.Initial reporter phone number exceeded allowable field length, initial reporter phone number is as follows: (b)(6).
|
|