• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF UNIT "ESG-400"; ELECTROSURGICAL GENERATOR ESG-400

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF UNIT "ESG-400"; ELECTROSURGICAL GENERATOR ESG-400 Back to Search Results
Model Number WB91051W
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/25/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user¿s report was not confirmed.There were no error codes noted during the evaluation.If additional information becomes available following the device evaluation, a supplemental report will be filed.
 
Event Description
The customer reported that his olympus hf generator unit was producing an e433 error code during procedure preparation.Additional details relating to the patient and the event have been requested, but no response has been received at this time.There was no patient harm or consequence reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF UNIT "ESG-400"
Type of Device
ELECTROSURGICAL GENERATOR ESG-400
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18139900
MDR Text Key328171690
Report Number9610773-2023-03300
Device Sequence Number1
Product Code GEI
UDI-Device Identifier04042761076838
UDI-Public04042761076838
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K203682
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWB91051W
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/27/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-