• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON Back to Search Results
Model Number 97810
Device Problems Overheating of Device (1437); Communication or Transmission Problem (2896)
Patient Problems Burning Sensation (2146); Discomfort (2330); Insufficient Information (4580)
Event Date 09/18/2021
Event Type  malfunction  
Manufacturer Narrative
This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for urinary dysfunction/sacral nerve stim and gastrointestinal/ pelvic floor.It was reported that information was received from a patient (pt) regarding a recharger.The patient reported seeing a recharger not found message, the recharger would connect and then it would start beeping and lose connection and the power light on recharger turned orange.The issue started saturday.The patient reported discomfort while they were recharging on the call and said it is warm by the battery a little after they started the recharging, it was noted the site was not fully healed.The following troubleshooting steps were performed; reset the recharger, located the ins by looking for incision and/or palpating the ins, repositioned the recharger, moved the charger away from the lead, recommended using recharger over a thin layer of clothing.Suggested the patient sit down instead of standing.The troubleshooting steps that were taken on the call improved the issue, they feel it a little bit with the belt and clothing.Patient got 40% battery charge, excellent connection, and my therapy on.By end of call patient had reached 50% charge of stimulator battery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18139989
MDR Text Key328236679
Report Number3004209178-2023-21276
Device Sequence Number1
Product Code EZW
UDI-Device Identifier00763000203849
UDI-Public00763000203849
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/14/2022
Device Model Number97810
Device Catalogue Number97810
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/20/2021
Initial Date FDA Received11/15/2023
Date Device Manufactured04/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
"SEE H10...."
Patient Age61 YR
Patient SexFemale
-
-