ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/86
Device Problems Break (1069); Difficult to Remove (1528)
Patient Problem Fatigue (1849)
Event Date 09/22/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented for an implant procedure.During the procedure, it was noted that the stylet could not be pulled out from the left ventricular (lv) lead.Eventually, the stylet was pulled out along with the lv lead and the lead was broke.The lead was replaced during the procedure to resolve the issue.The patient experienced fatigue during the procedure, but remained in a stable in condition afterwards.
 
Manufacturer Narrative
The reported events of stylet could not be removed and ¿proximal tip had broken¿ were confirmed.Final analysis found that as received, a partial lead (connector portion) was returned in one piece with the stylet stuck inside the lead.Visual inspection of the lead found that the connector pin and connector cap were pulled out of the connector assembly along with the inner coil which is consistent with procedural damage and the polytetrafluoroethylene (ptfe) stylet coating was bunched up/clogged with the inner coil distal to the connector pin.The cause of the reported events was due to bunched up ptfe coating of the stylet that prevented the removal of the stylet and excessive forces resulted in the connector pin and cap to be pulled out of the connector assembly.A device history record review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18140414
MDR Text Key328172727
Report Number2017865-2023-52509
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeIZ
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberS000090715
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/07/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-