Catalog Number 5C4482
Device Problems Crack (1135); Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2023
Event Type  malfunction  
Manufacturer Narrative
E1: initial reporter address: (b)(6).The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the patient connector of a minicap extended life pd transfer set cracked and did not fit closely to the titanium adapter.This was observed prior to use of the devices for peritoneal dialysis therapy.The titanium adapter remained in use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h10.H10: the device was received for evaluation.A visual inspection with the naked eye noted damage on the patient connector duplicating the reported issue.Functional testing including leak, clear passage, and clamp function testing were performed with no issues noted.Integrity testing was performed between patient adapter and an in-lab titanium adapter; there was no connection issue or leak observed.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP TRANSFER SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18141390
MDR Text Key328183652
Report Number1416980-2023-05946
Device Sequence Number1
Product Code KDJ
UDI-Device Identifier00085412007731
UDI-Public(01)00085412007731
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K152675
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number5C4482
Device Lot NumberH22K16057
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/16/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
N/A.
-
-