Model Number V60
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/04/2023
Event Type  malfunction  
Event Description
Philips received a complaint from the service engineer, reporting that the v60 ventilator had a battery that would not charge.There was no patient involvement when the issue occurred.No patient or user harm reported.A philips field service engineer (fse) evaluated the device and observed that the battery would not charge.The fse then replaced the battery to resolve the issue.
 
Manufacturer Narrative
H10: the suspected battery was returned to philips, and the failure investigation was performed.The technician reported the following conclusion/root cause, "product investigation lab (pil) is able to confirm the complaint.The battery indicates that it is at 100% capacity, therefore unable to charge, while at a lower voltage than the minimum acceptance criteria.Battery also failed to provide power after removing ac (alternating current) power.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18142028
MDR Text Key328189089
Report Number2518422-2023-30441
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/09/2023
Initial Date FDA Received11/15/2023
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-