Clearcorrect findings: patient was on step 1 when the symptoms started.Patient had difficulty breathing and a burning sensation.Patient has not been formally diagnosed with any allergies, but stated they had similar symptoms from other products such as acrylic nails and certain hair dyes.The device was rinsed before use with water and appeared to be clean, the reaction was not noticed until the patient was at home.Patient removed the devices after 3-4 hours of use with burning and itchy gingiva symptoms.The symptoms started to relieve after the aligners were removed.Photos were not taken of the reaction.A review of the treatment plan shows no issues that would contribute to patient discomfort.Case ledger indicates the devices were shipped on august 16, 2023.Clinical evaluation: the patient reported swollen gums, burning sensation as well as difficulty breathing when step 1 of the aligner series was worn.The symptoms were relieved after the aligners were discontinued and not worn.Unfortunately no clinical details regarding the description of the gums, locations and photographs were provided.A review of the treatment plan did not reveal any issues that could contribute to the patient's symptoms.Although the data are somewhat limited, there is sufficient information to suggest that allergic phenomenon may be responsible for the patient's symptoms.It is noted that the patient exhibited similar symptoms to hair dye and acrylic nails.
|