BECTON, DICKINSON AND COMPANY (BD) SYR ABG PRESET 3(1.6) LL 23X1 CE; BLOOD SPECIMEN COLLECTION DEVICE
|
Back to Search Results |
|
Catalog Number 364327 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/23/2023 |
Event Type
malfunction
|
Event Description
|
It was reported when using the, syr abg preset 3(1.6) ll 23x1 ce, blood leaked out of the syringe.No patient impact.The following information was provided by the initial reporter: "when the child was given arterial blood gas, it was found that the arterial blood had spread to the lumen through the piston.".
|
|
Manufacturer Narrative
|
H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
|
Manufacturer Narrative
|
H.6.Investigation summary: material #: 364314.Lot/batch #: 3075469.Bd had not received samples or photos for investigation.Therefore, 10 retention samples from bd inventory were evaluated by visual examination and functional testing, each drawn with water, and no issues were observed relating to leakage as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode leakage.Bd was not able to identify a root cause for the indicated failure mode.
|
|
Event Description
|
It was reported when using the, syr abg preset 3(1.6) ll 23x1 ce, blood leaked out of the syringe.No patient impact.The following information was provided by the initial reporter: "when the child was given arterial blood gas, it was found that the arterial blood had spread to the lumen through the piston.".
|
|