Section e3: occupation is patient/consumer the coaguchek xs test strip lot number was 68883822 with an expiration date of 30-sep-2024.The coaguchek xs test strip lot number was 66283721 with an expiration date of 30-jun-2024.All strip lots include the updated labeling and states: "caution: the coaguchek xs meter is pre-set to display results in the international normalized ratio (inr).The meter is also capable of displaying results in seconds (displayed on meter as sec) and % quick (displayed on meter as %q) (a measuring unit used mainly by healthcare professionals in europe).When testing for inr, you must confirm that the measured result is displayed in inr prior to using the result and whenever the date and time settings are modified.To reset the measuring unit to inr, follow the instructions in the testing a blood sample section of the coaguchek xs system user manual for self-testing, version 8.0 and higher or call the roche diagnostics technical service center at 1-800-428-4674 for assistance, if the result is displayed in %q or sec." on a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.The meter and test strips were requested for investigation.
|