Updated fields: b4, d9, g3, g6, h2, h3, h4, h6 (type of investigation, investigation findings, component codes, investigation conclusions), h10, h11.Corrected fields: e1, e3, h6 (medical device - problem code).A getinge field service engineer (fse) evaluated the unit and confirmed that the printer intermittently did not print the numerical data after the ecg strip.The fse replaced the recorder, which resolved the issue.All calibration, functional and safety tests were performed.The unit was returned to the customer and cleared for clinical use.The following investigation was performed by technician of the maquet failure analysis and testing dept.(fat) wayne, nj.The failure analysis and testing dept.Received recorder, with a reported unit failure of intermittent printing.The fat performed a visual inspection and found the part to be in good condition.The fat installed the part into cs300 test fixture and tested the part to factory specifications per the cs300 service manual.It was found that the part would not print properly.Fat was able to verify the reported issue.
|