The mayfield modified skull clamp (a1059) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - the investigation of the returned device found no device deficiencies that would have contributed to the reported complaint.Service & repair (s&r) could not duplicate slippage.There was minimal movement in the swivel lock assembly and a residue buildup; however, it is unlikely this contributed to a slippage.No issues could be found with the ratchet extension and the torque screw assembly.To resolve all issues, the internal parts to adjust the unit were replaced according to the manufacturer specifications and the skull clamp's swivel lock assembly was cleaned.After completing the reassembly, the skull clamp was subjected to a successful function test.New components were added to replace worn internal parts, and general maintenance and cleaning were performed.Root cause - the complaint is not confirmed.Probable root cause of the head movement is improper or suboptimal positioning of the patient in the skull clamp.No further investigation is required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|