Mdr report# mw5146670 was received with the following information:
"an 8cm head laceration caused by three point mayfield device originally positioned by surgical team.Laceration on left side of head, sutured at end of case using 3-0 biosin.Attending notified, charge nurse notified.Patient head slipped lower during case.Resident asked to secure head fellow.".
|
The mayfield skull clamp was not returned for evaluation and lot number information has not been provided; therefore, an evaluation of the device could not be performed, the device history record (dhr) could not be reviewed, and the definite root cause cannot be identified.Based on the reported complaint, probable root cause is improper or suboptimal placement of the skull clamp on the patient.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|