It was reported that during a patient infusion of parenteral nutrition, the solution set leaked.It was further reported that the set was connected to a pump, and when the pump door was opened, it was identified that the set was damaged and loaded incorrectly.According to the reporter, the infused volume indicated by the pump did not match the remaining solution in the parenteral nutrition bag.The infusion was stopped, and the parenteral nutrition was withdrawn.As a result of the event, the patient developed hypoglycemia and received additional treatment of 50% dextrose.No additional information is available.
|
H10: the actual device was not available; however, ten (10) retained samples were evaluated.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.The reported condition was not verified on the retained samples.The most probable cause of the condition is due to the misuse of the equipment by the end user.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|