• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK XS PT TEST; PROTHROMBIN TIME TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK XS PT TEST; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 04625358016
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/13/2023
Event Type  malfunction  
Manufacturer Narrative
The coaguchek inrange meter serial number was (b)(6).Section e3: occupation is patient/consumer.The meter and test strips were requested for investigation per the reporter's request.The product has not been received at this time.If the product is returned in the future, a follow-up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Per product labeling, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods.".
 
Event Description
The initial reporter complained of discrepant low inr results with a coaguchek inrange meter compared to an unknown laboratory method.On (b)(6) 2023 the result from the meter was 2.2 inr.The result from the laboratory was 2.89 inr.On (b)(6) 2023 the result from the meter was 2.3 inr.The result from the laboratory was 3.68 inr.On (b)(6) 2023 the result from the meter was 2.4 inr.The result from the laboratory was 3.73 inr.The timeframe between the meter and laboratory results was not provided.The patient¿s therapeutic range was not provided.
 
Manufacturer Narrative
The meter and test strips were received for investigation.The returned test strips were measured with the returned meter with a high-level control sample: testing results (qc range: 2.6 - 3.2 inr): qc 1: 2.9 inr.Qc 2: 3.0 inr.Qc 3: 2.9 inr.All inr values were within the specified target ranges, confirming the functionality of the coaguchek measuring system.No error messages occurred.The returned material and the retention material meet the specifications.The investigation did not identify a product problem.The cause of the event could not be determined.Section d9 was updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS PT TEST
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18153893
MDR Text Key328398414
Report Number1823260-2023-03640
Device Sequence Number1
Product Code GJS
Combination Product (y/n)Y
Reporter Country CodeIT
PMA/PMN Number
K170960
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number04625358016
Device Lot Number67810013
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/16/2023
Supplement Dates Manufacturer Received11/23/2023
Supplement Dates FDA Received12/01/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-