BOSTON SCIENTIFIC CORPORATION MOMENTUM X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) Back to Search Results
Model Number G138
Device Problems False Positive Result (1227); Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Inaccurate Synchronization (1609)
Patient Problem Electric Shock (2554)
Event Date 11/01/2023
Event Type  malfunction  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) potentially delivered inappropriate therapy for an atrial driven rhythm.Technical services (ts) reviewed the reports and noted that anti-tachycardia pacing (atp) and shock therapy was delivered, and the rhythm accelerated.A shock converted the rhythm.Additionally, ts stated that there was a pacemaker mediated tachycardia (pmt) episode that appeared to be atrial driven.Also, there was a signal artifact monitor (sam) episode due to atrial fibrillation (af) on the atrial channel that resulted in a vector switch.The device remains in use.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM X4 CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18154239
MDR Text Key328327025
Report Number2124215-2023-64815
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589188
UDI-Public00802526589188
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/30/2022
Device Model NumberG138
Device Catalogue NumberG138
Device Lot Number386854
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/11/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age49 YR
Patient SexMale
-
-