Model Number DW ARTHROSCOPY FLUID MANAGEMENT DEV |
Device Problem
No Display/Image (1183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/05/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that there is a loose connection from time to time with tubing (ar-6420).The device has also turned black during surgery.There was no harm for patient, operator or third party.Update swit (b)(6) 2022: the surgery could be finished successfully with the same pump and tubing.
|
|
Manufacturer Narrative
|
The complaint allegation was not confirmed.One ar-6480 dualwave arthroscopy fluid management system was returned for investigation.The returned device was visually inspected and no apparent damage was noted.Further review of the pump sub-assembly and components showed no issues with the latch door or tubing connector.The returned device was assembled with a new ar-6410 arthroscopy pump tubing and was tested and evaluated under normal use conditions to see if the issue(s) reported could be reproduced.The pump was powered on and functioned as intended with no error messages and no audible alarms triggered.The complaint tube set was not sent for evaluation.A clamp test was performed as defined in the user guide (dfu-0212 rev 1 ¿ section 5.2) to simulate an over pressure failure on the allege pump, and to verify if an error message and/or audible alarm will be triggered.The results of the clamp test confirmed the pump triggered an alarm and provided an error as intended/expected.The pump operated under normal conditions for two hours.Review of complaint records for serial number n13840f18 showed that the device has no previous complaints.
|
|
Search Alerts/Recalls
|