Model Number MC2AVR1-DELSYS |
Device Problems
Leak/Splash (1354); Activation, Positioning or Separation Problem (2906); Positioning Problem (3009)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/05/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that intra-operative while flushing the delivery system of the leadless implantable pulse generator (ipg), fluid was coming out of the handle where the deflection button was located.It was reported that the delivery system was hard to deflect after the ipg was retrieved, and that the deployment button was sluggish.The ipg system was attempted/not used. no patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product event summary: the full delivery system was returned and analyzed.The delivery system tether lock housing leaks.The articulation of the delivery system was out of plane.Fluid pathway leak was observed on the delivery system.The delivery system outer shaft was kinked/buckled.The delivery system inner shaft was mechanically kinked/bent.The analyst noted the full delivery system was returned with the device intact in the device cup.The outer shaft was kink/buckled at 5.5 cm from the distal end of the delivery system.The inner shaft was kinked at 1.5 cm and 2 cm from the distal end of the recapture cone.There was a leak while flushing within the delivery system handle between the tether lock housing the tether lock.There were no anomalies found with the deployment button.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|