Other text: d10.Device available for evaluation, h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.D3, g1, 2 email is: regulatory.Responses@icumed.Com.One device was received.Per visual inspection, cracked tank cover, noisy pump, stained reservoir, cracked and leaking enclosure, bent micro switch, and corroded drain fitting.Outdated printed circuit board (pcb) and power switch.Per functional testing, the pump was intermittent, and the temperature was unstable confirming both issues.The root cause for the pump not working properly and consistently was caused by the bent micro switch which triggers the pump.Not holding the temperature was caused by the pot on the pcb (used for temperature settings) being damaged.It was unknown what caused the condition.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It was deemed beyond economical repair and was scrapped.
|