• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA CRT-D QUAD RF HV; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA CRT-D QUAD RF HV; NO MATCH Back to Search Results
Model Number CD3365-40Q
Device Problem Use of Device Problem (1670)
Patient Problem Discomfort (2330)
Event Date 04/03/2023
Event Type  Injury  
Event Description
It was reported that a patient was experiencing pain around the top of his implantable cardioverter defibrillator (icd).The icd was explanted and replaced.The patient condition was not known.
 
Manufacturer Narrative
The device was at elective replacement indicator (eri) when received.A longevity calculation was performed and found the battery depletion was normal based on the device usage.The device was tested in the laboratory; telemetry, impedance, sensing, pacing, and high voltage (hv) output functions of the device were found to be normal.No anomaly was detected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA CRT-D QUAD RF HV
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18165849
MDR Text Key328477855
Report Number2017865-2023-72546
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508292
UDI-Public05414734508292
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2017
Device Model NumberCD3365-40Q
Device Lot NumberA000003980
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2023
Was the Report Sent to FDA? No
Date Manufacturer Received02/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/28/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA RV LEAD.; QUARTET LV LEAD.; TENDRIL RA LEAD.
Patient Outcome(s) Required Intervention;
Patient Age82 YR
-
-