The manufacturer was contacted in reference to innospire elegance device.The manufacturer received information alleging the unit was hot and had a burning smell while in use.There is no allegation of serious or permanent harm or injury.No medical intervention was required by the patient.No warping, melting, voids/gaps, smoke, or flames were reported.The device was returned to the manufacturer, and during evaluation the technician was unable to confirm the thermal event.The device was visually inspected and no signs of thermal or other events were visible on the plastic parts of the device.The npc filter was inserted and dirty.The nebulizer was connected to test bench where power was applied and the device power's on, fan spins, but pump engages, light amount of medicine dispersing indicating little to no pressure.Light environmental contamination (the inside of the device appears to have been recently cleaned): dirt, dust, and debris attached to the fan and pump assembly.No noticeable thermal or damage inside the device.Upon disassembly of the unit verifying only low pressure at 1.4 psi.There is evidence of lubricant leakage, internal pump-contamination, and piston-seal failure.Over time contamination will eventually cause the motor to overheat, seals to fail, premature eol issues, or eventually total mechanical seizure.The root cause of the complaint was determined to be possible contamination and/or a ventilation issue.The wear and tear may have been caused from contamination, and/or using with/without a dirty npc filter installed.This degraded performance is linked to internal motor/pump contamination that is being sucked inside through the npc air inlet port during each use.Eventually it wears the seals, lowering performance, and causing early end of life (eol) issues to the device.
|