Catalog Number 329491 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/23/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
E.1.Initial reporter phone #: (b)(6).H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the bd nano¿ 2nd gen pen needle was leaking.The following was translated from chinese to english: the patient complained of injecting insulin according to the method taught by the doctor.After injecting insulin according to the previous method, there was no reaction after the needle was removed.Now that the insulin needle has been replaced.When the needle is pulled out, liquid will be exposed from the needle.
|
|
Event Description
|
It was reported that the bd nano¿ 2nd gen pen needle was leaking.The following information was provided by the initial reporter, translated from chinese to english: the patient complained of injecting insulin according to the method taught by the doctor.After injecting insulin according to the previous method, there was no reaction after the needle was removed.Now that the insulin needle has been replaced.When the needle is pulled out, liquid will be exposed from the needle.
|
|
Manufacturer Narrative
|
H.6.Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.H3 other text : see h.10.
|
|