A field service engineer (fse) was at the customer¿s site to address the reported event.Fse was able to replicate the error by performing a shutdown with both substrate and alcohol, but error persists.Upon observation during the shutdown, fse noticed dust/lint was in the substrate line.The fse disassembled the 3-way solenoid and found dust/lint inside the solenoid valve causing the error.Fse removed the dust/lint from the substrate line by purging the substrate line and replaced the 3-way solenoid valve.Fse validated the analyzer by performing multiple shutdowns, daily checks, calibration, and quality control run without error and results within acceptable range.No further action required by field service.The aia-360 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(6).There were no similar complaints identified during the searched period.Aia-360 training manual under analyzer error codes, section: troubleshooting: (2017) substrate purge failure cause: no substrate was dispensed at daily maintenance.Solution: check substrate level and replace as necessary, repeat daily maintenance.The most probable cause of the reported event was due to the dirty 3-way solenoid valve.
|