• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Other text: b3: date of event and d4: udi section are unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that in the operating room, during a procedure the device had electrical issues.The customer stated "intermittent issue potentially".No patient or clinical injury reported.
 
Manufacturer Narrative
Other, other text: additional information is provided in g.1., h.2, h.3., and h.6.Device evaluation: one device was received.A visual inspection found a scratched front cover, discolored line cord, cracked water tank, corroded quick connect and the stand off was missing behind the lcd.During the functional testing, the customer reported problem was unable to be duplicated but other problems were found.The lcd was missing a standoff and the water tank leaked.The pcb was replaced as well as the front cover and global display label, line cord, water tank cover, and quick connect.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.For all enquiries or follow-up questions related to the record, do not use regulatory.Responses@smiths-medical.Com located in sections g.1., please direct those to the following: regulatory.Responses@icumed.Com.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key18172272
MDR Text Key328638959
Report Number3012307300-2023-10630
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-390
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? No
Date Returned to Manufacturer11/08/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/31/2023
Initial Date FDA Received11/20/2023
Supplement Dates Manufacturer Received04/04/2024
Supplement Dates FDA Received04/30/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/16/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-