Model Number G156
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439); Failure to Read Input Signal (1581); High Capture Threshold (3266)
Patient Problem Bradycardia (1751)
Event Date 11/10/2023
Event Type  malfunction  
Event Description
It was reported that the patient with cardiac resynchronization therapy defibrillator (crt-d) device exhibited out of range amplitude.Upon review, there were no noise and impedance measurements were stable.Myopotential testing was performed, and no low r waves were seen.The patient had some premature ventricular contraction (pvc).The non-boston scientific right atrial (ra) lead thresholds were noted.It was noted that when the ra threshold testing was performed, the device stopped pacing.However further review of latitude showed some v episodes and atrial tachy response (atr) episodes with noise on all, likely due to electromagnetic interference (emi).The patient underwent heart surgery few months back and there was significant dropped in the p waves.The plant was to perform an x-ray and further discuss with the physician.This device and non-boston scientific leads remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN X4 CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATORS (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18173123
MDR Text Key328630584
Report Number2124215-2023-65413
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534669
UDI-Public00802526534669
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 11/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/10/2020
Device Model NumberG156
Device Catalogue NumberG156
Device Lot Number379236
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/13/2023
Initial Date FDA Received11/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/27/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age76 YR