MICRO THERAPEUTICS, INC. DBA EV3 HYPERGLIDE; CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY
|
Back to Search Results |
|
Model Number 104-4127 |
Device Problem
Burst Container or Vessel (1074)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/18/2023 |
Event Type
malfunction
|
Event Description
|
Medtronic received a report that the hyperglide balloon was prepared as indicated in the instructions for use (ifu).During inflation inside the neurovascular anatomy, the balloon got ruptured.No patient symptoms or further complications were reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received that the physician tested the balloon prior to use and the balloon performed as intended during testing.The guidewire tip was advanced 5cm out of the catheter tip during inflation.It was noted that the balloon got ruptured throughout.Contrast ratio was 50/50.The physician did not shape the guidewire tip.The balloon was inflated and deflated 2 times.The injection rate was steady.A 3cc syringe was used during inflation/deflation.
|
|
Search Alerts/Recalls
|
|
|