Model Number MI1200 SYNCHRONY PIN |
Device Problems
Impedance Problem (2950); Insufficient Information (3190)
|
Patient Problem
Failure of Implant (1924)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Event Description
|
The user suddenly stopped hearing with the device at the beginning of november 2023.
|
|
Manufacturer Narrative
|
Additional information: according to the currently available information, damage to the active electrode as might be caused by an external mechanical impact appears likely.However, to determine an exact root cause a device investigation of the explanted device is necessary.The concerned device was explanted but has not been received for investigation yet.
|
|
Event Description
|
The user suddenly stopped hearing with the device at the beginning of (b)(6) 2023.The user was explanted.
|
|