MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH SYNCHRONY; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH SYNCHRONY; COCHLEAR IMPLANT Back to Search Results
Model Number MI1200 SYNCHRONY PIN
Device Problems Impedance Problem (2950); Insufficient Information (3190)
Patient Problem Failure of Implant (1924)
Event Type  malfunction  
Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event Description
The user suddenly stopped hearing with the device at the beginning of november 2023.
 
Manufacturer Narrative
Additional information: according to the currently available information, damage to the active electrode as might be caused by an external mechanical impact appears likely.However, to determine an exact root cause a device investigation of the explanted device is necessary.The concerned device was explanted but has not been received for investigation yet.
 
Event Description
The user suddenly stopped hearing with the device at the beginning of (b)(6) 2023.The user was explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRONY
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key18181866
MDR Text Key328661905
Report Number9710014-2023-01044
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09008737093969
UDI-Public(01)09008737093969
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMI1200 SYNCHRONY PIN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/13/2023
Date Device Manufactured10/05/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age5 YR
Patient SexFemale
-
-