• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR Back to Search Results
Model Number 97716
Device Problems Overheating of Device (1437); Use of Device Problem (1670); Charging Problem (2892); Communication or Transmission Problem (2896)
Patient Problem Burning Sensation (2146)
Event Date 12/25/2021
Event Type  malfunction  
Manufacturer Narrative
Continuation of d10: product id 97755 lot# serial# (b)(6) implanted: explanted: product type recharger section d information references the main component of the system.Other relevant device(s) are: product id: 97755, serial/lot #: (b)(6), ubd: , udi#: (b)(4) h3: analysis of the 97755 intellis recharger (s/n (b)(6)) revealed that "no device found" message was shown.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient (pt) regarding an external device.The reason for call was since about 3 days ago, pt mentioned their recharger antenna cord was getting hot to the touch when they charged their implanted neurostimulator( ins); pt said they had to hold their recharger antenna coil in "exactly the right position" to get their recharger antenna to connect to their ins to charge.Pt said the entered passive recharge mode at one point in time and the middle number changed from 0 to 90 "percent" when they moved the recharger antenna very slightly.Pt said "it was like a isometric exercise" to get the ins to charge.During the call, the pt turned the relay box over to read the serial number and the recharger antenna disconnected from the ins while charging the ins.Recharger antenna was very sensitive.Ins charge was "low." pt said their program they used was working well, but they needed to charge their ins every other day.No symptoms or patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18187153
MDR Text Key328977857
Report Number3004209178-2023-23122
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169781740
UDI-Public00643169781740
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Notification
Type of Report Initial
Report Date 11/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2020
Device Model Number97716
Device Catalogue Number97716
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2022
Initial Date Manufacturer Received 12/28/2021
Initial Date FDA Received11/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1535-2021
Patient Sequence Number1
Treatment
"SEE H10...."
Patient Age64 YR
Patient SexMale
-
-