• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: UNKNOWN ROTAREX; THROMBECTOMY & ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNKNOWN ROTAREX; THROMBECTOMY & ATHERECTOMY Back to Search Results
Catalog Number 80223
Device Problem Detachment of Device or Device Component (2907)
Patient Problems Extravasation (1842); Perforation (2001); Perforation of Vessels (2135)
Event Date 10/26/2023
Event Type  Injury  
Event Description
It was reported that during a recanalization procedure in the right iliac artery, the patient allegedly experienced perforation of the external iliac artery and had extravasation of the right external iliac artery.It was further reported that the wire was allegedly withdrawn from the patient by the catheter and caused the gold tip of the wire to uncoil.Reportedly, the covered stenting of the right external iliac artery was performed to treat extravasation.The procedure was completed using another device.The patient is reported to be stable now.
 
Manufacturer Narrative
H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the rotarex products that are cleared in the us.The pro code and 510 k number for the rotarex products are identified in d2 and g4.H10: as the lot number for the device was not provided, a review of the device history records could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that during a recanalization procedure in the right iliac artery, the patient allegedly experienced perforation of the external iliac artery and had extravasation of the right external iliac artery.It was further reported that the wire was allegedly withdrawn from the patient by the catheter and caused the gold tip of the wire to uncoil.Reportedly, the covered stenting of the right external iliac artery was performed to treat extravasation.The procedure was completed using another device.The patient is reported to be stable now.
 
Manufacturer Narrative
H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the rotarex products that are cleared in the us.The pro code and 510 k number for the rotarex products are identified in d2 and g4.H10: manufacturing review: as the lot number for the device was not provided, a review of the device history records could not be performed.Investigation summary: the sample was not returned for evaluation and a physical investigation was not possible.Also, no images or videos were provided for review.The user reported damage of the guidewire, due to sample reported malfunction cannot be confirmed.Therefore, the investigation is inconclusive for the reported issue.A clear root cause could not be identified but a damaged guidewire represents a known inherent risk.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H11: h6 (patient).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAREX
Type of Device
THROMBECTOMY & ATHERECTOMY
Manufacturer (Section D)
UNKNOWN
BR 
Manufacturer (Section G)
UNKNOWN
BR  
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key18190585
MDR Text Key328781182
Report Number3008439199-2023-00198
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K211738
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number80223
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/27/2023
Initial Date FDA Received11/22/2023
Supplement Dates Manufacturer Received02/05/2024
Supplement Dates FDA Received02/14/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-