ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. TEGO¿ CONNECTOR; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number D1000
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/20/2023
Event Type  malfunction  
Event Description
The event involved a tego¿ connector where it was reported that while being connected to a patient where the end of the tego cap peeled causing high arterial pressures upon initiation of hemodialysis.Unexpected or prolonged care is unknown and invasive procedure is not provided or no applicable.There was no medical intervention was required.
 
Manufacturer Narrative
The device was returned for evaluation; however, testing has not yet been completed.
 
Manufacturer Narrative
The used d1000 tego connector was visually inspected for occlusion that would result in resistance to flow during use.The fluid path was completely open without evidence of occlusion.A subsequent fluid test showed the fluid flowed without restriction.Finally the used d1000 tego connector was pressure and vacuum leak tested without failure.The complaint was unable to be replicated or confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEGO¿ CONNECTOR
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18190762
MDR Text Key328783424
Report Number9617594-2023-01018
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K053106
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD1000
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/24/2023
Initial Date FDA Received11/22/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-