STRYKER MEDICAL LONDON DBA CHG HOSPITAL BEDS SPIRIT SELECT; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Catalog Number 5700000000
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/10/2023
Event Type  malfunction  
Event Description
It was reported that the device had exposed bare wires.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Manufacturer Narrative
This device was originally reported as unknown.The model number was identified, and the record was updated to reflect that.The initial report listed 0001831750-medical as the fda registration number.However, the proper fda registration number, associated with this model number, should be 3006433555-medical (londonchgbed).A visual and functional inspection was performed by stryker field service technician.It was found that the unit had exposed bare wires (potential for ac shock) due to broken/damaged power cord assembly.
 
Event Description
It was reported that the device had exposed bare wires.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPIRIT SELECT
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
STRYKER MEDICAL LONDON DBA CHG HOSPITAL BEDS
1020 adelaide st. s.
portage MI N6E 1 R6
CA  N6E 1R6
Manufacturer (Section G)
STRYKER MEDICAL LONDON DBA CHG HOSPITAL BEDS
1020 adelaide st. s.
portage MI N6E 1 R6
CA   N6E 1R6
Manufacturer Contact
marleca lane
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key18192761
MDR Text Key328812810
Report Number0001831750-2023-01348
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number5700000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received11/10/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-