This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Based on the results of the investigation, the suggested events were confirmed.Phenomenon (1): the phenomenon, ¿power of the device is turned off¿, was not reproduced, and the root cause could not be identified; phenomenon (2): the phenomenon, ¿error code e03¿, occurred due to faulty main board.However, the root cause could not be identified.Replacement of the main board was recommended.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 8 years since the subject device was manufactured.Olympus will continue to monitor field performance for this device.
|