|
Model Number CDHFA500Q |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that the patient presented in clinic for generator changeout procedure.During procedure, it was found that the new implantable cardioverter defibrillator (icd) header set screw on the right ventricular port was unable to engage with the lead.The icd was not implanted and an alternate near at hand was implanted instead on (b)(6) 2023.Patient condition was stable.
|
|
Manufacturer Narrative
|
The reported field event set screw anomaly could not be confirmed.The device was returned in one piece for analysis.A malfunction based on analysis was found.Visual inspection of the header attachment area detected discoloration.All set screws were able to latch and secure test leads normally.Telemetry, impedance, sensing, pacing, and high voltage (hv) output functions of the device were normal with no other anomalies found.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
|
|
Manufacturer Narrative
|
The reported field event set screw anomaly could not be confirmed.The device was returned in one piece for analysis.A malfunction based on analysis was found.Visual inspection of the header attachment area detected discoloration.All set screws were able to latch and secure test leads normally.Telemetry, impedance, sensing, pacing, and high voltage (hv) output functions of the device were normal with no other anomalies found.
|
|
Search Alerts/Recalls
|
|
|