Catalog Number P310NUS |
Device Problems
Gas Output Problem (1266); Tidal Volume Fluctuations (1634)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Other text: b3: date of event is unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the "oxygen was 75 to 81 percent air mix to high failed test on level 1 service sheet 4.2.8".Unit fails when testing o2 at tidal volume 125 at 30 bpm.Occurred during preventive maintenance.There was no patient involvement and no patient or clinical injury.The outcome of event was ongoing.
|
|
Manufacturer Narrative
|
One sample was received for evaluation.Visual inspection revealed damage on the input label surrounding the patient outlet connector and a minor bend on the front panel above the manometer.No issues found during the testing of the device.Service history review identified there was no indication that the complaint was related to a service of the device within the review period.For all enquiries or follow-up questions related to the record, do not use (b)(4) located in sections g.1., please direct those to the following: (b)(4).
|
|